CS301 Features

CS301 Simulator
  • ECG waveforms – VF, VT fast, VT slow, VT poly, AFIB, AFLTR, SVT, S TACH, NSR, ASYS, SINUS PVC, S BRDY, J BRDY, 2nd I, 2ND II PVC, 2nd II, 3rd.
  • Convert – simulate cardioversion by activating convert* feature.
  • Defib discharge - indicator illuminates for two seconds when simulator is shocked by a defib set to 50J, or more.
  • Pacer pulse - paced beat is displayed and indicator flashes when simulator is paced by an external pacer. Capture occurs when pacer current is set at 60mA to 70mA, or more.
  • Connectors – RA, LA, LL, RL and PADS
  • Battery – powered by one 9 volt alkaline battery. Low battery indicator illuminates when battery needs to be replaced. Battery saver feature powers off simulator automatically when not in use. Battery life is approximately 25 hours.
  • Three year warranty covers material and workmanship.

Convert

Convert automatically from a running rhythm to a waiting rhythm when simulator is shocked by a defib set to 50J, or more.

  1. Press the convert key to engage. The indicator of the running rhythm will pulse brighter.
  2. Press the key of the rhythm you wish to activate after the defib discharge. This is the waiting rhythm. The indicator of the waiting rhythm will flash on and off.
  3. Shock simulator with defib set to 50J or more. The waiting rhythm becomes the new running rhythm.
  4. To cancel the convert operation before a shock is delivered, press the convert key again, or press the key of the running rhythm. The convert operation cancels automatically after two minutes if a shock is not delivered.

Operating Instructions [PDF]

Simulator Base Number: 9801-


PADS Connectors

Click on an image for a larger view.

Physio-Control/Medtronic (-001) -001

Philips / HeartStart (-008) -008

ZOLL (-000) -000

Welch Allyn/R2 (-017) -017

Marquette (-002) -002